Результаты исследований: Вклад в журнал › Статья › Рецензирование
Результаты исследований: Вклад в журнал › Статья › Рецензирование
}
TY - JOUR
T1 - INFLUENCE OF THERMOBEAM PROCESSING ON THERMOLUMINESCENCE AND OPTICALLY STIMULATED LUMINESCENCE PROPERTIES OF THE SOLID-STATE DETECTOR OF IONISING RADIATION ON THE BASIS OF ANION-DEFECTIVE CORUNDUM
AU - Litovchenko, E.N.
AU - Soloviev, S.V.
AU - Milman, I.I.
AU - Moiseikin, E.V.
AU - Surdo, A.I.
PY - 2012
Y1 - 2012
N2 - In this paper TL and OSL-properties of detectors TLD-500K subjected to thermobeam processing in the temperature range from 50 to 800 °С are resulted. It is shown that increase in the intensity of TL and OSL signal in the peak is connected with intensive filling of the traps competing to basic - dosimetric traps at irradiation temperature of 300- 400 °С, and with change in concentration of the basic luminescence centers.
AB - In this paper TL and OSL-properties of detectors TLD-500K subjected to thermobeam processing in the temperature range from 50 to 800 °С are resulted. It is shown that increase in the intensity of TL and OSL signal in the peak is connected with intensive filling of the traps competing to basic - dosimetric traps at irradiation temperature of 300- 400 °С, and with change in concentration of the basic luminescence centers.
UR - https://elibrary.ru/item.asp?id=19015563
M3 - Article
VL - 55
SP - 105
EP - 109
JO - Известия вузов. Физика
JF - Известия вузов. Физика
SN - 0021-3411
IS - 11-3
ER -