Standard

Accuracy of a Bellows vacuum meter with an LC-generator. / Korolev, I.; Alekseenko, N.; Porodnov, B.
в: Measurement Techniques, Том 49, № 9, 2006, стр. 899-906.

Результаты исследований: Вклад в журналСтатьяРецензирование

Harvard

APA

Vancouver

Korolev I, Alekseenko N, Porodnov B. Accuracy of a Bellows vacuum meter with an LC-generator. Measurement Techniques. 2006;49(9):899-906. doi: 10.1007/s11018-006-0209-2

Author

Korolev, I. ; Alekseenko, N. ; Porodnov, B. / Accuracy of a Bellows vacuum meter with an LC-generator. в: Measurement Techniques. 2006 ; Том 49, № 9. стр. 899-906.

BibTeX

@article{af5c30284d7840169b0291c75e46b62f,
title = "Accuracy of a Bellows vacuum meter with an LC-generator",
abstract = "Errors of pressure measurement are considered. The error as a function of the pressure for a vacuum gauge rated at up to 1000 Pa is calculated as an example. It is 0.15 ≤ δp/p ≤ 0.2% over a pressure range of 14 ≤ p ≤ 1000 Pa and it increases hyperbolically up to 100% with p = 0.01 Pa. The parametric reliability of the instrument and its immunity to an accidental increase in pressure the atmospheric level is substantiated. {\textcopyright} Springer Science+Business Media, Inc. 2006.",
author = "I. Korolev and N. Alekseenko and B. Porodnov",
year = "2006",
doi = "10.1007/s11018-006-0209-2",
language = "English",
volume = "49",
pages = "899--906",
journal = "Measurement Techniques",
issn = "0543-1972",
publisher = "Springer",
number = "9",

}

RIS

TY - JOUR

T1 - Accuracy of a Bellows vacuum meter with an LC-generator

AU - Korolev, I.

AU - Alekseenko, N.

AU - Porodnov, B.

PY - 2006

Y1 - 2006

N2 - Errors of pressure measurement are considered. The error as a function of the pressure for a vacuum gauge rated at up to 1000 Pa is calculated as an example. It is 0.15 ≤ δp/p ≤ 0.2% over a pressure range of 14 ≤ p ≤ 1000 Pa and it increases hyperbolically up to 100% with p = 0.01 Pa. The parametric reliability of the instrument and its immunity to an accidental increase in pressure the atmospheric level is substantiated. © Springer Science+Business Media, Inc. 2006.

AB - Errors of pressure measurement are considered. The error as a function of the pressure for a vacuum gauge rated at up to 1000 Pa is calculated as an example. It is 0.15 ≤ δp/p ≤ 0.2% over a pressure range of 14 ≤ p ≤ 1000 Pa and it increases hyperbolically up to 100% with p = 0.01 Pa. The parametric reliability of the instrument and its immunity to an accidental increase in pressure the atmospheric level is substantiated. © Springer Science+Business Media, Inc. 2006.

UR - http://www.scopus.com/inward/record.url?partnerID=8YFLogxK&scp=33751538348

U2 - 10.1007/s11018-006-0209-2

DO - 10.1007/s11018-006-0209-2

M3 - Article

VL - 49

SP - 899

EP - 906

JO - Measurement Techniques

JF - Measurement Techniques

SN - 0543-1972

IS - 9

ER -

ID: 42416228